201
201
These sensors are designed for temperature measurement in liquid and gaseous non-corrosive media, and also corrosive media that do not damage protective cover material. Sensors are recommended to be used together with UNKJ thermowell.
201 modification sensors can be of explosion protection type 0ExiallCT6 X or 1ExdllCT6 according to GOST 30852.10-2002. 202 modification sensors can only be of explosion protection type 0ExiallCT6 X. Further details see in "Alternative designs" below.
PRTD 201K and 202K thermometers are designed for cryogenic temperature measurement. PRTD 201H and 202H is an alternative to copper thermocouple for temperature measurement up to 200°С. Fitting standard size ¾ on Customer's request.
Terminal heads can be equipped with measuring transmitters with 4-20 mА DC output signal according to GOST 26.011 and/or digital signal via HART, PROFIBUS-PA, FOUNDATION Fieldbus protocol, and also cable glands for additional cable fastening, and metal hose fastening, if required.
02 modification sensors can be equipped with remote transmitters.
Connection diagram |
2-wire |
tolerance class В, С (see item 9 at page 6-5) |
|
3-wire |
tolerance class АА, А, В, С |
||
4-wire |
|||
Vibration resistance GOST R 52931 |
group V3 |
||
Nominal (rated) |
6.3 MPa |
||
Seismic resistance MSK-64 |
grade 9 for unit installation up to 70 m above zero mark |
||
Climatic category GOST 15150 |
UHL2Ambient temperature: |
-60..+120ºС for general purpose products |
|
-60..+85 ºС for Ex version with analogous signal |
|||
-55..+85ºС for products with unified output signal |
|||
Calibration |
- GOST 8. 461-2009 without measuring transmitters; |
Group of operating |
Probability of |
Specified lifetime |
Mean |
Guarantee |
I |
0.95 for 40 000 hours |
5 years |
10 years |
5 years |
II |
0.95 for 16 000 hours |
2 years |
4 years (6 years)* |
2 years |
III |
0.95 for 8 000 hours |
1 year |
2 years |
1 year |
* - Increased mean lifetime with 0,6 failure-free operation possibility for the given period.
Measurement accuracy of sensors with 4-20 mА DC signal and/or digital signal via HART protocol
Output signal |
Nominal |
Measurement |
Output |
Nominal |
Measurement |
4-20mА |
AA3T25; А3T25 |
0,25 % · tn or 0,5 °С |
4-20mА |
AAxH25, AxH25 |
0,25 % · tn or 0,3 °С |
B3T70 |
0,7 % · tn or 1,0 °С |
AxH10, BxH10 |
0,1 % · tn or 0,15 °С |
||
A3T40 |
0,4 % · tn or 0,5 °С |
BxH70 |
0,7 % · tn or 1,0 °С |
* - tn thermocouple range should be multiplied by specified value in в %. Select higher value.
** - «х» means number of wires in resistance thermometer connection diagram, х=3 or 4. For example, АА4Н25 or B3H70.
TC type |
Thermowell diameter, mm |
Thermowell material |
Modification |
Group of |
Tolerance |
Application |
Cal. int. |
Mean lifetime |
CRTD |
10, 8, 6 |
S10 |
201, 202 |
II |
A |
– 50…+120 |
2 years |
4 years |
10 |
S13 |
B, C |
– 50…+200 |
|||||
PRTD |
10, 8, 6 |
S10 |
201Н,202Н |
I |
A, B, С |
– 50…+200 |
5 years |
10 years |
II |
AA |
– 50…+150 |
2 years |
4 years |
||||
III |
– 50…+250 |
1 year |
2 years |
|||||
201К, 202К, |
II |
A, B, С |
– 196…+200 |
2 years |
4 years |
|||
10 |
S13 |
|||||||
201, 202 |
I |
– 50…+300 |
5 years |
10 years |
||||
II |
– 50…450 |
2 years |
4 years |
|||||
III |
В, С |
– 50…600 |
1 years |
2 years |
Response time:
Sensor response time depending on diameter, s |
|
d = 6; 8 |
d = 10 |
16 |
20 |
PRTD |
|
Exi |
201 |
G |
— |
А |
21 |
— |
2х |
P100 |
— |
В |
3 |
Н10 |
— |
S10 |
— |
8 |
— |
L |
/ |
l |
|
G1/2 |
1 |
|
2 |
3 |
4 |
|
5 |
6 |
|
7 |
8 |
|
9 |
10 |
11 |
|
12 |
|
13 |
|
14 |
|
15 |
|
16 |
NN |
Name |
Code |
Description |
|||||||
1 |
Sensor type |
CRTD |
Copper resistance temperature detector |
|||||||
PRTD |
Platinum resistance temperature detector |
|||||||||
2 |
Explosion protection |
Not filled |
general purpose electrical equipment |
|||||||
Exi |
0ExiaIICT6 X, intrinsically safe circuit |
|||||||||
Exd |
1ExdIICT6, explosion-proof enclosure |
|||||||||
3 |
Modification |
201, 202 |
see drafts and operating temperature |
|||||||
4 |
Version (see "operating temperature" table) |
Not filled |
Standard version from -50 to +600°С |
|||||||
К |
Cryogenic version from -200 to +200°С |
|||||||||
G |
Low-temperature version from -50 to +200°С |
|||||||||
5 |
Wire terminal head |
0 |
Free electrode ends |
50, 60 Modification wires |
||||||
2 |
miniature plug |
|||||||||
4 |
standard plug |
|||||||||
Cable gland |
0 |
standard cable gland |
Terminal heads |
|||||||
А |
for unarmoured cable in metal hose RZCH-15 |
Terminal heads ≥ 14 |
||||||||
В |
for unarmoured cable in metal hose MRPI-15 |
|||||||||
C |
for unarmoured cable in metal hose DN18 |
|||||||||
D |
for unarmoured cable in metal hose DN20 |
|||||||||
E |
for unarmoured cable in metal hose DN12 |
|||||||||
F |
for pipe assembly, М20х1,5 internal thread |
|||||||||
G |
for pipe assembly, output external thread G1/2 |
|||||||||
H |
for unarmoured cable, 6.5 - 14 mm diameter |
|||||||||
I |
for unarmoured cable, 3.2 - 8.7 mm diameter |
|||||||||
J |
for cable with internal/external sheath diameter: 6,1-11,7/9,5-15,9, single-line wire armoured cable |
|||||||||
6 |
Sensor terminal |
10 |
plastic head |
IP55 |
general purpose |
|||||
15; 16;17; 18; 19 |
aluminium head |
IP66/IP68 |
1ExdIICT6 / 0ExiaIICT6 X |
|||||||
20; 22 |
aluminium head |
IP65 |
general purpose |
|||||||
14; 21; 23; 24; 25; 26; 28; 29 |
aluminium head |
IP66 |
0ExiaIICT6 Х or general purpose |
|||||||
27 |
stainless steel |
IP66 |
0ExiaIICT6Х or general purpose |
|||||||
50 |
silicon / shield / silicon |
IP65 |
0ExiaIICT6 or general purpose |
|||||||
60 |
PTFE / shield / PTFE |
|||||||||
7 |
Number of SE |
Not filled |
one sensitive element |
|||||||
2хPt100 |
two sensitive elements |
|||||||||
8 |
Type of TC |
50М, 100М, 50P, 100P, Pt100, Pt500, Pt1000 |
NSC according to GOST 6651-2009 |
|||||||
9 |
Tolerance class |
АА, А, В, С |
Tolerance class according to GOST 6651-2009 |
|||||||
10 |
Connection diagram |
3, 4 |
3- and 4-wire connection diagram |
for АА, А class |
||||||
2, 3, 4 |
2-, 3-, and 4-wire connection diagram |
for В, С class |
||||||||
11 |
Output signal, transmitter accuracy symbol, |
Not filled |
analogous signal (Ohm) according to NSC |
|||||||
T25; T25 |
For terminal |
4-20 mА |
for AA3 and A3 |
|||||||
T40 |
for A3 |
|||||||||
T70 |
for B3 |
|||||||||
H10, H10 |
4-20 mА +HART |
for А4, В4, А3, В3 |
individual |
|||||||
H25, H25 |
for АА4, А4, АА3, А3 |
|||||||||
H70 |
for В3, В4 |
|||||||||
12 |
Outer cable |
S10 |
steel 12Cr18Ni10T |
|||||||
S13 |
steel 10Cr17Ni3M2T |
|||||||||
13 |
Outer diameter |
6, 8 |
size (mm) on Customer's choice. |
S10 |
||||||
10 |
S10, S13 |
|||||||||
14 |
Probe length, L |
50÷3150 |
probe length L to the working end, mm |
|||||||
15 |
Extension neck length ℓ |
Not filled |
if there is no extension |
If terminal head is selected |
||||||
30÷500 |
specify size (mm) if there is an extension |
|||||||||
Extension wire length ℓ |
100÷30,000 |
specify dimensions in mm 500, 1000, 2000 3150 and more |
If wire |
|||||||
16 |
Fitting standard size |
Not filled |
if threaded fitting M20x1.5 is used |
|||||||
Specify thread size |
in other cases |
PRTD Exi 201–A21 – Pt100 – А4Н10 – S10 – 8 – 250/100 |
|||
Product type |
PRTD |
Platinum resistance temperature detector |
|
Explosion protection |
Exi |
0ExiaIICT6 X |
|
Modification |
101 |
with welded fitting |
|
Cable gland |
А |
for H3ZC DN15 |
|
Wring (terminal head code) |
21 |
IP66 |
|
Type of TC |
Pt100 |
|
|
Tolerance class |
А |
class A |
|
Connection diagram |
4 |
4-wire |
|
Output signal (accuracy class) |
Н10 |
4-20mА+HART |
|
Sheath material |
S10 |
steel 12Cr18Ni10T |
|
Working end diameter |
8 |
mm |
|
Probe length L |
250 |
mm |
|
Extension neck length ℓ |
100 |
mm |
|
PRTD Exi 201-028–Pt100-В3Н70 – S13 - 10 – 800 |
|||
Product type |
PRTD |
Platinum resistance temperature detector |
|
Explosion protection |
Exi |
0ExiaIICT6 X |
|
Modification |
201 |
without extension |
|
Cable gland |
0 |
standard |
|
Wring (terminal head code) |
28 |
IP66 |
|
Type of TC |
Pt100 |
|
|
tolerance class and connection diagram |
В3 |
Class B, 3 wire diagram |
|
Output signal (accuracy class) |
Н70 |
4-20mА+HART |
|
Sheath material |
S13 |
steel 10Cr17Ni3M2T |
|
Working end diameter |
10 |
mm |
|
Probe length L |
800 |
mm |
|
without extension |
― |
|
|
PRTD 201Н-010–100P-В3 – S10 - 8 – 10/120 |
|||
Product type |
PRTD |
Platinum resistance temperature detector |
|
Explosion protection |
|
general purpose |
|
Modification |
101Н |
low-temperature |
|
Cable gland |
― |
standard |
|
Wring (terminal head code) |
10 |
IP55 |
|
Type of TC |
100P |
|
|
Tolerance class |
В |
class B |
|
Connection diagram |
3 |
3-wire |
|
Output signal (accuracy class) |
― |
analogous |
|
Sheath material |
S10 |
steel 12Cr18Ni10T |
|
Working end diameter |
8 |
mm |
|
Probe length L |
100 |
mm |
|
Extension neck length ℓ |
120 |
120 mm |
|
PRTD 201К-020–100P-В3 – S10 - 8 – 150/120 |
|||
Product type |
PRTD |
Platinum resistance temperature detector |
|
Explosion protection |
― |
general purpose |
|
Modification |
101К |
cryogenic |
|
Cable gland |
― |
standard |
|
Wring (terminal head code) |
20 |
IP65 |
|
Type of TC |
100P |
|
|
Tolerance class |
В |
class B |
|
Connection diagram |
3 |
3-wire |
|
Output signal (accuracy class) |
― |
analogous |
|
Sheath material |
S10 |
steel 12Cr18Ni10T |
|
Working end diameter |
8 |
mm |
|
Probe length L |
150 |
mm |
|
Extension neck length ℓ |
120 |
120 mm |