104

is designed for indoor air temperature measurement. Sensors can be of explosion protection type 0ExiaIICT6 X or 1ExdIICT6 according to GOST 30852.10-2002. For further details see "Alternative designs & Ordering information" below. Terminal heads of PRTD (CRTD) 104-014… and PRTD (CRTD) 104-018…  can be equipped with measuring transmitters with 4-20 mА DC output signal according to GOST 26.011 and/or digital signal via HART, PROFIBUS-PA, FOUNDATION Fieldbusprotocol, and also cable glands for additional cable fastening, and metal hose fastening, if required.

Technical characteristics

Connection diagram

2-wire

tolerance class B, C

3-wire

tolerance class АА, А, В, С

4-wire

Vibration resistance GOST R 52931

group V3

Seismic resistance MSK-64

grade 9 for unit installation up to 70 m above zero mark

Climatic category GOST 15150

UHL2Ambient temperature:

-60..+120ºС for general purpose products

-60..+85 ºС for Ex version with analogous signal

-55..+85ºС for products with unified output signal

Calibration

- GOST 8. 461-2009 for sensors without measuring transmitters;
- MP RT 2026-2013 – for sensors with installed measuring transmitters.

Response time

does not exceed 16 seconds

Reliability indicators

Group of operating
conditions

Probability of
failure-free operation

Specified lifetime

Mean
lifetime

Guarantee
conditions

I

0.95 for 40 000 hours

5 years

10 years

5 years

II

0.95 for 16 000 hours

2 years

4 years (6 years)*

2 years

* - Increased mean lifetime with 0,6 failure-free operation possibility for the given period.

Error limits

Measurement accuracy of sensors with 4-20 mА DC signal according to GOST 26.011 and/or digital signal via HART protocol 


Output
signal

Nominal
designation

Measurement
accuracy
limits

Output
signal

Nominal
designation**

Measurement
accuracy
limits

4-20mА

AA3T25; А3T25

0,25 % · tn  or 0,5 °С

4-20mА
+
HART

AAxH25, AxH25

0,25 % · tn or 0,3 °С

B3T70

0,7 % · tn  or 1,0 °С

AxH10, BxH10

0,1 % · tn or 0,15 °С

A3T40

0,4 % · tn  or 0,5 °С

BxH70

0,7 % · tn or 1,0 °С

*   - tn thermocouple range should be multiplied by specified value in в %. Select higher value.
**  - «х» means number of wires in resistance thermometer connection diagram, х=3 or 4. For example, АА4Н25 or B3H70.

Temperature range

TC type

Thermowell diameter, mm

Thermowell material

Group of    
operating  
conditions

Tolerance
class

Application
temperature, °С

Cal. int.

Mean lifetime

CRTD

6

S10

II

A, B, С

–50…+120 for general purpose sensors

2 years

4 years

PRTD

II

AA

2 years

4 years

–50…+85 for EX versions of
sensors with transmitters

I

A, B, С

5 years

10 years

Execution options

PRTD

 

Exi

104

А

21

Pt100

В

3

Н10

S10

6

L

1

 

2

3

 

4

5

 

6

 

7

8

9

 

10

 

11

 

12

 

NN

Name

Code

Description

1

Sensor type

CRTD

Copper resistance temperature detector

PRTD

Platinum resistance temperature detector

2

Explosion protection

Not filled

general purpose electrical equipment

Exi

0ExiaIICT6 X, intrinsically safe circuit

Exd

1ExdIICT6, explosion-proof enclosure

3

Modification

104

see drafts

4

Cable

0, А, В, С…

see cable gland description at page 6-10

5

Sensor terminal
head (see "Alternative
designs"
page 1-10)

13

plastic
head

IP55

general purpose

18

aluminium
head

IP66/IP68

1ExdIICT6 / 0ExiaIICT6 X

14

aluminium
head

IP66

0ExiaIICT6 X or
general purpose

6

Type of TC

50М, 100М, 50P, 100P, Pt100, Pt500, Pt1000

NSC according to GOST 6651-2009

7

Tolerance class

АА, А, В, С

Tolerance class according to GOST 6651-2009

8

Connection diagram

3, 4

3 or 4 diagram
is not valid for
013 head

3- and 4-wire connection diagram for class АА, А

2, 3, 4

2-, 3-, 4-wire connection diagram for class В, С

9

Output signal, measuring transmitter accuracy designation - see table 3 at page 6-3

Not filled

analogous signal (Ohm) according to NSC

T25; T25

For terminal
heads 14, 18

4-20 mА

for AA3 and A3

T40

for A3

T70

for B3

H10, H10

4-20 mА +HART

for АА4, А4, АА3, А3

individual
sensor graduation

H25, H25

for А4, В4, А3, В3,

H70

for В3, В4

10

Outer sheath material

S10

steel 12Cr18Ni10T

11

Outer diameter

6

dimensions in mm

12

Probe length, L

60÷200

probe length L to the working end, mm

Examples of product designation

PRTD 104–013 –100М – В2 – S10 – 6 – 100

Product type

CRTD

Copper resistance temperature detector

Explosion protection

general purpose

Modification

104

 

Cable gland

0

standard cable gland

Wring (terminal head code)

13

IP55

Type of TC

100М

 

tolerance class and connection diagram

В

class B

Connection diagram

2

2-wire

Output signal (accuracy class)

according to NSC

Sheath material

S10

steel 12Cr18Ni10T

Working end diameter

6

mm

Probe length L

100

mm

PRTD Exi 104–A14 – Pt100 – А4Н10 – S10 – 8 – 80

Product type

PRTD

Platinum resistance temperature detector

Explosion protection

Exi

0ExiaIICT6 X

Modification

104

 

Cable gland

А

for metal hose H3ZC DN15

Wring (terminal head code)

14

IP66

Type of TC

Pt100

 

Tolerance class

А

class A

Connection diagram

4

4-wire

Output signal (accuracy class)

Н10

4-20mА+HART

Sheath material

S10

steel 12Cr18Ni10T

Working end diameter

6

mm

Probe length L

80