02.12

These sensors are designed for temperature measurement in continuous metal cast mold.
Mounting element is a spring-loaded mounting fitting with М6 or М8х1 thread, or a bayonet screw BS12 (see page 2-54). Mounting element type should be specified in order.
02.12 thermocouple design is covered by utility model patents No. 79667.

Technical characteristics

Hot junction

one

Ungrounded, Grounded, Exposed

Vibration resistance GOST R 52931

group V3

Nominal (rated) pressure

0,1 MPa

Calibration

MI 3090-2007 - for sensor fitting length 20 to 250 mm
GOST 8.338-2001 - for sensor fitting length larger than 250 mm

Response time τ 0,63:

ungrounded

0,3 seconds    ( τ 0,95=0,6 s)

grounded

0,25 seconds    ( τ 0,95=0,5 s)

exposed

0,1 seconds    ( τ 0,95=0,4 s)

Temperature range

CT type

Group of operating conditions

Application
temperature, ºС

Calibration interval / specified lifetime

Failure-free operation

Mean lifetime

CTK

II

-40…+200

2 years

0.95 for 40 000 hours

4 years

III

-40…+900

1 year

0.95 for 16 000 hours

2 years

Execution options

CTK

 

02.12

0

60

c1

U

 

Т310

2

L

/

lc

/

l

-

 

1

 

2

 

3

4

 

5

 

6

7

 

8

 

9

 

10

 

11

 

12

 

13

 

NN

Parameter name

Code

Description

1

Sensor type

CTK

Mineral insulated cable thermocouple with type according to IEC 60584 and ASTM E230

2

Modification

02.12

 

3

Terminal head

0

Free electrode ends 50mm

IP65

general purpose

4

Wire version (see "Alternative designs" table, pages 1-14)

60

Stranded conductor, PTFE / shield / PTFE insulation - shield is insulated from sensor body and does not have a terminal

61

Stranded conductor, PTFE / shield / PTFE insulation - shield is insulated from sensor body and has a separate terminal

62

Stranded conductor, PTFE / shield / PTFE insulation - shield is connected with sensor body and has a separate terminal

5

Tolerance class

c0, c1;   c2

For further details see table 5 page 2-9

6

Hot junction design

G

grounded junction,

О

exposed junction,

Operation group only III

U

ungrounded junction

7

Number of wire couples

Not filled

1 wire couple

8

Cable outer sheath material

Т310, Т600

AISI 310S, Inconel 600

9

Outer diameter

2

mm

10

Probe length, L

50÷500

For 02.05 probe length can be up to 1250 mm

11

Wire length

250÷10,000

specify size (mm):250, 320, 500, 1000, 2000, 3150 and more

12

Cable part length ℓК

Not filled

70 mm

50÷1000

mm

13

Mounting element

М6, М8

Threaded fitting М6х1, М8х1

BS 12

Bayonet screw BS12